Theranos Investor Conference Presentation Deck
Lipids on miniLab Correlate to
Comparator Method for Venous Plasma
miniLab
(mg/dL)
miniLab
(mg/dL)
350
300-
250-
200-
150-
100
250-
200
150
100-
50
y = 0.97 x + 3.74
= 0.99
n = 103
100 150 200 250 300 350
Siemens ADVIA 1800 (mg/dL)
Total Cholesterol
Median bias = -0.9 %
ATE = + 9%
Points within ATE = 100%
y = 0.98 x + 4.01
2=0.95
n=103
LDL Cholesterol
Median bias = 2.4%
ATE = ± 12%
Points within ATE = 92%
50 100
150 200 250
Siemens ADVIA 1800 (mg/dL)
This technology has not been cleared or approved by the FDA and is not for sale in the United States.
miniLab
(mg/dL)
800-
600-
400-
200
0
0
140-
120-
miniLab 100-
(mg/dL)
80-
60-
40-
20-
y= 1.03 x 1.58
r² = 0.99
n = 103
200 400 600 800
Siemens ADVIA 1800 (mg/dL)
y = 1.03 x + 3.09
²=0.95
n-103
Legend:
Unity (y = x)
- Allowable total error
20
NCEP (Clin Chem 1988;34:193-201)
NCEP (Clin Chem 1997;43:2164-2168)
Triglycerides
Median bias = 0.6%
ATE = ± 15%
Points within ATE = 98%
HDL Cholesterol
Median bias = 9%
ATE = ± 13%
Points within ATE = 76%
40 60 80 100
140
Siemens ADVIA 1800 (mg/dL)
theranos
28View entire presentation